Anslutning, dräneringsledning:
6 x 8 mm (tubing must not exceed 4 feet and must drain straight down)
Anslutning, provledning:
4 x 6 mm stainless steel tubing
Antal kanaler:
1
Detektionsgräns:
Drift is negligible; 1 ppb
Elektriska certifieringar:
EMC
CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Safety
CAN/CSA C22.2 No. 61010-1
cETLus safety mark for:
- General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1
Flöde:
166 - 250 mL/min (10 - 15 L/h) recommended
Handboksspråk:
Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
Kommunikation (optional):
Analog:
Five 0-20 mA or 4-20 mA analog outputs on each analog output module
Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.
Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Kommunikationsmöjligheter:
Claros mA Output
Lagringsförhållanden:
-20 - 70 °C, 0 - 95% relative humidity, non-condensing
Material:
Counter-Electrode: Stainless steel; Working Electrode: Platinum
Matningskrav (Spänning):
24 V DC
Mätområde:
0 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable
ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL.
0 - 500 ppb hydrazine; programmable
0 - 500 ppb hydrazine; programmable
Omgivningstemperatur:
5 - 45 °C at 0 - 95% relative humidity (non-condensing)
Pausläge:
With 24 VDC
Programvara tillgänglig:
Cloud (Claros-enabled)
Provkrav:
Sample needs to be free of undissolved matter.
Provtemperatur:
5 - 45 °C
Reagens:
Sample conditioning with diisopropylamine, ammonia, or diethylamine
Reference system:
Ag/AgCl electrode with electrolyte KCl 0.1 M
Reläer:
Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Relativ luftfuktighet:
10 - 90% relative humidity (non-condensing)
Relay: Operational mode:
Primary or secondary measurement or timer/scheduler
Relay functions:
Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Repeterbarhet:
± 2% or 1 ppb (whichever is greater)
Responstid :
< 60 s
Security levels:
2
Tryckområde:
0.5 - 6 bar (7.2 - 87 psi) or 12 L/h
Underhållsintervall:
Monthly: Calibration and reagent refill
Vad ingår i leveransen:
Controller installed in Panel, Working Electrode, Reference Electrode, Counter Electrode, Basic User Manual
Vikt :
14.6 kg