Analoga utgångar:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal provströmmar:
1, 2, 4 or 8
Certifieringar :
CE compliant / UL certified
Cykeltid:
20 - 30 minutes
40 minutes (+ Total Hardness Ca)
Demineralised water:
For rinsing
Detektionsgräns:
Calcium Hardness: ≤ 10 mg/L
Total Alkalinity: ≤ 20 mg/L
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimension (H x B x D):
690 mm x 465 mm x 330 mm
Dränering:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garanti :
24 månader
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Kalibrering:
N.A.
Larm:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Mätmetod:
Acid-base titration (alkalinity only); Calmagite EDTA titration with LED dipping probe (combined)
Matningsspänning:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Mätområde:
Ca: 10 - 100 mg/L CaCO₃; Alk: 20 - 500 mg/L CaCO₃
Calcium Hardness: 10 - 100 mg/L CaCO₃; Total Alkalinity: 20 - 500 mg/L CaCO₃
Metodnamn:
Titrimetric
Omgivningstemperatur:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter:
Calcium Hardness; Total Alkalinity
Precision:
Better than 2% full scale range for standard test solutions
Prov: tryck:
By external overflow vessel
Provflödeshastighet:
100 - 300 mL/min
Provkvalitet:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Provtemperatur:
10 - 30 °C
Reagent Requirements:
Keep between 10 - 30 °C
Skyddsklass:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Störningar:
Hardness: Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter also may interfere with the end point. Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Alkalinity: Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Validation:
Automatic; frequency freely programmable
Vikt :
25 kg