Analoga utgångar:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal provströmmar:
1, 2, 4, or 8
Automatisk rengöring:
Ja
Certifieringar :
CE compliant / ETL certified
Cykeltid:
10 min (dilution + 5 min)
Demineralised water:
For rinsing / dilution
Detektionsgräns:
≤ 20 µg/L
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimension (H x B x D):
690 mm x 465 mm x 330 mm
Dränering:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garanti :
24 månader
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Kalibrering:
Automatic, 2-point; frequency freely programmable
Larm:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Mätmetod:
Colorimetric measurement using vanadate yellow method (450 nm)
Matningsspänning:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Mätområde:
0.1 - 10 mg/L PO₄-P
0.1 - 10 mg/L PO₄-P
Optional:
0.02 - 1 mg/L
0.05 - 2.5 mg/L
0.05 - 5 mg/L
0.8 - 40 mg/L (with internal dilution)
1.6 - 80 mg/L (with internal dilution)
10 - 1000 mg/L (with internal dilution)
Modell:
EZ1031
Omgivningstemperatur:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter:
Orthophosphate (PO₄-P)
Pausläge:
100 - 240 VAC, 50/60 Hz
Precision:
Better than 3% full scale range for standard test solutions
Prov: tryck:
By external overflow vessel
Provflödeshastighet:
100 - 300 mL/min
Provkvalitet:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Provtemperatur:
5 - 30 °C
Reagent Requirements:
Keep between 10 - 30 °C
Skyddsklass:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Störningar:
Positive interference is caused by Silica Arsenate if the sample is heated. Negative interferences are caused by Arsenate, Fluoride, Thorium, Bismuth, Sulphide, Thiosulphate, Thiocyanate or excess of Molybdate. Blue colour is caused by Ferrous Iron but this does not affect results for Ferrous Iron concentrations < 100 mg/L. If Nitric Acid is used, Chloride interferes from 75 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Vad ingår i leveransen:
EZ1031 Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 1 x empty 2.5L Reagent Container with Fittings (for Colour Solution)
Validation:
Automatic; frequency freely programmable
Vikt :
25 kg