Arbetsluftfuktighet:
Up to 90 % at 40 °C (104 °F) maximum
Beskrivning:
CL17 Chlorine Analyzer with Reagents for Total Chlorine
Cykeltid:
2.5 min
Detektionsgräns:
0.03 mg/L (ppm)
Display :
LCD, 3.5 inch digit measurement readout and six-character alphanumeric scrolling text line
Garanti :
24 månader
Inlopp:
¼" OD polyethylene tube, quick-disconnect fitting
Inloppstryck:
1 to 5 psi (0.07 - 0.34 bar), 1.5 psi (0.103 bar) is optimum
Kapsling:
ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Kommunikationsmöjligheter:
4-20 mA output
Larm:
Two alarms selectable for sample concentration alarm, analyser system warning, or analyser system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Leveransomfattning:
CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Ljuskälla:
Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life
Luftspolning:
0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psi (ca.1.4 bar) with ¼" OD tube, quick disconnect fitting
Matningskrav (Hz):
50/60 Hz
Matningskrav (Spänning):
100 - 115/230 V AC
Mätområde:
0 - 5 mg/L Cl₂
Monteringsform:
Wall mount
Noggrannhet:
± 5 % or ±0.03 mg/L (ppm) as CL2, whichever is greater
Omgivningstemperatur:
5 - 40 °C
Parameter:
Chlorine, Free & Total
Power requirements (Amps):
2,5 A
Pressure Regulator Included?:
No
Provflödeshastighet:
200 - 500 mL per minute minimum required
Provtemperatur:
5 - 40 °C (41 - 104 °F)
Repeterbarhet:
0.01 ppm eller 3.0 % , beroende på vilket som är störst
Skrivarutgångar:
One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground.
Störningar:
Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference:
1 mg/L Cr6+ = approximately 0.02 mg/L as Cl2.
Hardness must not exceed 1,000 mg/L CaCO3.
Tillbehör:
Ingen
Tillbehör som ingår:
Ingen
Tryckområde:
Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Vad ingår i leveransen:
CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Vikt :
11,3 kg