9184sc Total Free Chlorine (TFC) Amperometric Sensor

9184sc Total Free Chlorine (TFC) Amperometric Sensor
Produktnr: LXV432.99.00001
Pris per st Kontakta oss
Skickas inom 1 vecka

Reagentless analysis of Free Chlorine

Provides a wide continuous Chlorine measurement range: 0 - 20 mg/L

Integral temperature and pH sensors provide more accurate readings

Optional acidification (for cleaning) and intemittent flow accessories link in series so they only take up one sensor port controller

Compatible with multi-sensor, multi-parameter digital controllers

Vad finns i förpackningen

9184sc Chlorine Sensor Panel with pH Sensor, Membranes and Electrolyte and Manual.

Specifikationer

Beskrivning: 9184sc Total Free Chlorine Amperometric Sensor
Detektionsgräns: 0.005 mg/L HOCl
Flöde: 14 L/h (200 - 250 mL/min) auto-regulated by flow through cell
Garanti : 24 månader
Kalibreringsintervall : 2 months
Kalibreringsmetod : Zero calibration: electrically (automatic) or with de-chlorinated water
Calibration: comparison of lab method with process sample
pH calibration: single- or two-point using standards or comparison of lab method with process sample
Kapslingsklass: IP66 / NEMA 4X
Lagringsförhållanden: -20 °C - 60 °C
Mätintervall: Continuous
Mätområde: 0 - 20 mg/L HOCl
Mätprincip: Amperometric/Membrane (electrode, membrane, electrolyte)
Monteringsform: Flat, vertical surface (panel, stand, etc.)
Noggrannhet: 2% or ±10 ppb HOCl, whichever is greater at pH <8
  2% or ±10 ppb TFC, whichever is greater at pH <7.5
Omgivningstemperatur: 0 - 45 °C
Parameter: Total Free Chlorine: Sum of hypochlorous acid, hypochlorite and free chlorine
pH-område: 4 - 8 pH, acidification unit available for pH >8
Provintag: 0.25" outer diameter
Provtemperatur: 2 - 45 °C
Provutlopp: 0.5" inner diameter
Relativ luftfuktighet: 0 - 90 % non-condensing
Responstid : 90 % in less than 90 seconds
Standard deviation: 0.7 %
Störningar: No interference from chloramines.
Chlorine dioxide and ozone interfere with measurement.
Temperaturkompensation: Automatic over sample temperature range
  Automatic over sample temperature range
Tryckområde: 0.1 - 2 bar in flow cell
Underhållsintervall: Measurement cell: 6 months for membrane and electrolyte for typical application
pH: 1 - 1.5 years for typical application
Vad ingår i leveransen: 9184sc Chlorine Sensor Panel with pH Sensor, Membranes and Electrolyte and Manual.
Vikt : 6,5 kg